DRI-CHEM NX600 : Parameters and Specifications | Fujifilm [India]

In order to provide you with a better browsing experience and to improve our site functionality, we use cookies and other tracking technologies. Detailed information on the use of cookies on our site and how to opt out is provided in our Cookie Policy. By clicking into any content on this site, you consent that we can store and access cookies and other tracking technologies as described in our Cookie Policy.

India

DRI-CHEM NX600

An automated clinical chemistry analyzer with large, easy-to-operate touch panel and improved workflow for setting consumables with the Consumable Disk.  

The content on this page is intended to healthcare professionals and equivalents.

Parameters

Classification Parameter Measurement range *1 Measurement time
(min.)
Unit (A) Unit (B)
Biochemical tests Enzymes ALP 14 – 1183 U/L 0.23 – 19.76 μ Kat/L 4
AMYL 10 – 1800 U/L 0.17 – 30.06 μ Kat/L 5
CHE 5 – 500 U/L 0.08 – 8.35 μ Kat/L 4.5
CKMB 1 – 300 U/L 0.02 – 5.01 μ Kat/L 5
CPK 10 – 2000 U/L 0.17 – 33.40 μ Kat/L 4
GGT 10 – 1200 U/L 0.17 – 20.04 μ Kat/L 5
GOT/AST 10 – 1000 U/L 0.17 – 16.70 μ Kat/L 4
GPT/ALT 10 – 1000 U/L 0.17 – 16.70 μ Kat/L 4
LAP 10 – 500 U/L 0.17 – 8.35 μ Kat/L 4
LDH 50 – 900 U/L 0.84 – 15.03 μ Kat/L 2
LIP 20 – 1000 U/L 0.33 – 16.70 μ Kat/L 5
General chemistry ALB 1.0 – 6.0 g/dL 10 – 60 g/L 6
BUN 5.0 – 140.0 mg/dL 1.79 – 49.98 mmol/L 4
Ca 4.0 – 16.0 mg/dL 1.00 – 4.00 mmol/L 4
CRE 0.2 – 24.0 mg/dL 18 – 2122 μmol/L 5
DBIL 0.1 – 16.0 mg/dL 2 – 274 μmol/L 5
GLU 10 – 600 mg/dL 0.6 – 33.3 mmol/L 6
HDL-C 10 – 110 mg/dL 0.26 – 2.84 mmol/L 6
IP 0.5 – 15.0 mg/dL 0.16 – 4.84 mmol/L 5
Mg 0.2 – 7.0 mg/dL 0.08 – 2.88 mmol/L 4.5
NH3 10 – 500 μg /dL 7 – 357 μmol/L 2
TBIL 0.2 – 30.0 mg/dL 3 – 513 μmol/L 6
TCHO 50 – 450 mg/dL 1.29 – 11.64 mmol/L 6
TCO2 5 – 40 mmol/L 5 – 40 mmol/L 5
TG 10 – 500 mg/dL 0.11 – 5.65 mmol/L 4
TP 2.0 – 11.0 g/dL 20 – 110 g/L 6
UA 0.5 – 18.0 mg/dL 30 – 1071 μmol/L 4
Electrolytes Na 75 – 250 mEq/L 75 – 250 mmol/L 1
K 1.0 – 14.0 mEq/L 1.0 – 14.0 mmol/L
Cl 50 – 175 mEq/L 50 – 175 mmol/L
Immunological test CRP 0.3 – 7.0 mg/dL 3 – 70 mg/L 5
  • * There are parameters which may not be available in your area. For details please contact your local distributor.
  • *1 Unit (A) or (B) is available

Calculations

Calculated Parameter Indication Unit Equation
LDL Cholesterol LDL-C mg/dL LDL-C = TCHO value - (HDL-C vlaue + TG value/5)
mmol/L LDL-C = TCHO value - (HDL-C value + TG value/2.2)
non-HDL Cholesterol non-HDL-C mg/dL or mmol/L non-HDL = TCHO value - HDL-C value
Globulin GLOB g/dL or g/L GLOB = TP value - ALB value
Albumin/Globulin ratio ALB/GLOB - ALB/GLOB = ALB value / (TP value - ALB value)
BUN/Creatinine ratio BUN/CRE - BUN/CRE = BUN value / CRE value
GOT/GPT ratio (AST/ALT ratio) GOT/GPT (AST/ALT) - GOT/GPT=GOT value / GPT value (AST/ALT=AST value / ALT value)
Sodium/Potassium ratio Na/K - Na/K=Na value / K value
Anion Gap Anion Gap mEq/L or mmol/L Anion Gap = Na value - (Cl value + TCO2 value)

 

Main specifications

Measurement test

Colorimetry 28 tests, Electrolytes 3 tests

Throughput

Colorimetry 120 test/hour, Combined 128 test/hour

Number of sample rack

1

Number of incubator cell

Colorimetry 13, Electrolytes 1

Measurement time

Colorimetry 2 to 6 minutes/test, Electrolytes 1 minute/3 tests (Na-K-Cl)

Sample type

Plasma, Serum, Whole blood*2

Sample volume

Colorimetry 10μL/test, Electrolytes 50μL/3 tests (Na-K-Cl), CRP 5μL/test

Data transmission to PC

RS 232C (1 port), USB (2 port), LAN (1 port)

Data print

Thermal Printer

Electrical requirements

Single phase AC; 100 - 240 V ±10%; 50 to 60 Hz

Display

7-inch color touch panel

Dimensions

470 (W) ✕ 360 (D) ✕ 370 (H) mm

Weight

Approx. 28 kg (with PF unit)/26 kg (without PF unit)

Operating temperature

15 to 32˚C (59 to 89 °F)

Operating humidity

30 to 80%RH

Altitude

Up to 2,000 m (810 hPa)

  • *2 NH3-W: Whole blood only
  •     NH3-P: Plasma only
  •     Na-K-Cl: Plasma, Serum, Whole blood
  •     Other test items: Plasma, Serum

DRI-CHEM NX600 Series

  NX600 NX600i
Electrolyte tests
Plasma Filter Function -
Automatic dilution
  • * Please contact your local distributor for availability.

Option Items

FUJI DRI-CHEM OPERATOR ID EDITOR KIT

For registering and editing operator ID on a PC and importing them into an analyzer.

Appricable Sample /Tube

  • φ16 × 100 mm blood collection tube
  • φ13 × 100 mm blood collection tube
  • φ13 × 75 mm blood collection tube
  • 1.5 mL Fuji tube 
  • 0.5 mL Fuji tube
  • φ16 × 100 mm blood collection tube (when using PF)
  • φ13 × 100 mm blood collection tube (when using PF)
  • φ13 × 75 mm blood collection tube (when using PF)