The content on this page is intended to healthcare professionals and equivalents.
- Sensitivity: 94.3%, Specificity: 100%
- Reactive to all 4 serotypes: DEN-1,2,3,and 4
- Target specimen: Whole blood, serum, and plasma
- Time to result: Approximately 15 minutes
- Compatible with both visual and reader detection
The immunochromatographic assay was designed based on silver halide photography to amplify the size of gold colloidal particles approximately a 100-fold, leading to the improvement of detection sensitivity. This offers a greater ability to detect even in small amounts of antigens.
Correlation was compared with RT-PCR using 140 serum specimens.
Sensitivity: 94.3% (66/70, 95% CI: 85.8-98.2%)
Specificity: 100% (70/70, 95% CI: 93.8-100.0%)
RT-PCR | ||||
---|---|---|---|---|
Positive | Negative | Total | ||
FUJIFILM SILVAMP Dengue NS1 Ag | Positive | 66 | 0 | 66 |
Negative | 4 | 70 | 74 | |
Total | 70 | 70 | 140 |
- Test Cartridge × 10 tests
- R1 tube (including dilution buffer) × 10 × 0.45 mL/tube
- Pipette × 10 pieces
- Tube rack (for R1 tube) × 1 piece
- Nozzle (for R1 tube) × 10 pieces
- Instructions for use × 1 piece
- FUJIFILM SILVAMP Reader