Parameters

ClassificationParameterMeasurement range *1Measurement time
(min.)
Unit (A)Unit (B)
Biochemical testsEnzymesALP14 – 1183U/L0.23 – 19.76μ Kat/L4
AMYL10 – 1800U/L0.17 – 30.06μ Kat/L5
CHE5 – 500U/L0.08 – 8.35μ Kat/L4.5
CKMB1 – 300U/L0.02 – 5.01μ Kat/L5
CPK10 – 2000U/L0.17 – 33.40μ Kat/L4
GGT10 – 1200U/L0.17 – 20.04μ Kat/L5
GOT/AST10 – 1000U/L0.17 – 16.70μ Kat/L4
GPT/ALT10 – 1000U/L0.17 – 16.70μ Kat/L4
LAP10 – 500U/L0.17 – 8.35μ Kat/L4
LDH50 – 900U/L0.84 – 15.03μ Kat/L2
LIP20 – 1000U/L0.33 – 16.70μ Kat/L5
General chemistryALB1.0 – 6.0g/dL10 – 60g/L6
BUN5.0 – 140.0mg/dL1.79 – 49.98mmol/L4
Ca4.0 – 16.0mg/dL1.00 – 4.00mmol/L4
CRE0.2 – 24.0mg/dL18 – 2122μmol/L5
DBIL0.1 – 16.0mg/dL2 – 274μmol/L5
GLU10 – 600mg/dL0.6 – 33.3mmol/L6
HDL-C10 – 110mg/dL0.26 – 2.84mmol/L6
IP0.5 – 15.0mg/dL0.16 – 4.84mmol/L5
Mg0.2 – 7.0mg/dL0.08 – 2.88mmol/L4.5
NH310 – 500μg /dL7 – 357μmol/L2
TBIL0.2 – 30.0mg/dL3 – 513μmol/L6
TCHO50 – 450mg/dL1.29 – 11.64mmol/L6
TCO25 – 40mmol/L5 – 40mmol/L5
TG10 – 500mg/dL0.11 – 5.65mmol/L4
TP2.0 – 11.0g/dL20 – 110g/L6
UA0.5 – 18.0mg/dL30 – 1071μmol/L4
ElectrolytesNa75 – 250mEq/L75 – 250mmol/L1
K1.0 – 14.0mEq/L1.0 – 14.0mmol/L
Cl50 – 175mEq/L50 – 175mmol/L
Immunological testCRP0.3 – 7.0mg/dL3 – 70mg/L5
Comprehensive S-panelTP, ALB, ALP, GLU, TBIL, IP, TCHO, GGT, GPT/ALT, Ca, CRE, BUN
Small Screening S-panelALP, TBIL, GGT, GPT/ALT, CRE, GOT/AST
  • * There are parameters which may not be available in your area. For details please contact your local distributor.
  • *1 Unit (A) or (B) is available

Calculations

Calculated Parameter

Indication

Unit

Equation

LDL Cholesterol

LDL

mg/dL

LDL-C = TCHO value - (HDL-C vlaue + TG value/5)

mmol/L

LDL-C = TCHO value - (HDL-C value + TG value/2.2)

non-HDL Cholesterol

non-HDL

mg/dL or mmol/L

non-HDL = TCHO value - HDL-C value

Globulin

GLOB

g/dL or g/L

GLOB = TP value - ALB value

Albumin/Globulin ratio

ALB/GLOB

-

ALB/GLOB = ALB value / (TP value - ALB value)

BUN/Creatinine ratio

BUN/CRE

-

BUN/CRE = BUN value / CRE value

GOT/GPT ratio (AST/ALT ratio)

GOT/GPT (AST/ALT)

-

GOT/GPT=GOT value / GPT value (AST/ALT=AST value / ALT value)

Sodium/Potassium ratio

Na/K

-

Na/K=Na value / K value

Anion Gap

Anion Gap

mEq/L or mmol/L

Anion Gap = Na value - (Cl value + TCO2 value)

 

Main specications

Measurement test

Colorimetry 28 tests, Electrolytes 3 tests

Throughput

Colorimetry 180 test/hour, Combined 190 test/hour

Number of sample rack

5

Number of incubator cell

Colorimetry 13, Electrolytes 1

Measurement time

Colorimetry 2 to 6 minutes/test,

Electrolytes 1 minute/3 tests (Na-K-Cl)

Sample type

Plasma, Serum, Whole blood*2

Sample volume

Colorimetry 10μL/test,

Electrolytes 50μL/3 tests (Na-K-Cl), CRP 5μL/test

Data transmission to PC

RS 232C (1 port), USB (1 port), LAN (1 port)

Data print

Thermal Printer

Electrical requirements

Single phase AC; 100 - 240 V ±10%; 50 to 60 Hz

Display

7-inch color touch panel

Dimensions

500 (W) ✕ 380 (D) ✕ 410 (H) mm

Weight

Approx. 33kg

Operating temperature

15 to 32˚C (59 to 89F)

Operating humidity

30 to 80%RH

  • *2 NH3-W: Whole blood only
  •     NH3-P: Plasma only
  •     Na-K-Cl: Plasma, Serum, Whole blood
  •     Other test items: Plasma, Serum

DRI-CHEM NX700 Series

 NX700NX700i
Electrolyte tests
Plasma Filter Function-
Automatic dilution
  • *Please contact your local distributor for availability.

Option Items

Barcode Reader

Barcode reader is available as option item to read sample ID on sample tube.

Sample Rack

For φ16 × 100 mm blood collection tube
For φ13 × 100 mm blood collection tube
For φ13 × 75 mm blood collection tube
For 1.5 mL Fuji tube 
For 0.5 mL Fuji tube
For φ16 × 100 mm blood collection tube (when using PF)
For φ13 × 100 mm blood collection tube (when using PF)
For φ13 × 75 mm blood collection tube (when using PF)