You are accessing from the United States. To browse Fujifilm USA website, please click the following link.

Fujifilm USA Website

DRI-CHEM NX500 - Parameters and Specifications

A multi-purpose automatic dry-chemistry analyzer featuring remarkably short turnaround time, a touch screen interface and reliability.

Overview

The content on this page is intended to healthcare professionals and equivalents.

Parameters

ClassificationParameterMeasurement range *1Measurement time
(min.)
Unit (A)Unit (B)
Biochemical testsEnzymesALP14 – 1183U/L0.23 – 19.76μ Kat/L4
AMYL10 – 1800U/L0.17 – 30.06μ Kat/L5
CHE5 – 500U/L0.08 – 8.35μ Kat/L4.5
CKMB1 – 300U/L0.02 – 5.01μ Kat/L5
CPK10 – 2000U/L0.17 – 33.40μ Kat/L4
GGT10 – 1200U/L0.17 – 20.04μ Kat/L5
GOT/AST10 – 1000U/L0.17 – 16.70μ Kat/L4
GPT/ALT10 – 1000U/L0.17 – 16.70μ Kat/L4
LAP10 – 500U/L0.17 – 8.35μ Kat/L4
LDH50 – 900U/L0.84 – 15.03μ Kat/L2
LIP20 – 1000U/L0.33 – 16.70μ Kat/L5
General chemistryALB1.0 – 6.0g/dL10 – 60g/L6
BUN5.0 – 140.0mg/dL1.79 – 49.98mmol/L4
Ca4.0 – 16.0mg/dL1.00 – 4.00mmol/L4
CRE0.2 – 24.0mg/dL18 – 2122μmol/L5
DBIL0.1 – 16.0mg/dL2 – 274μmol/L5
GLU10 – 600mg/dL0.6 – 33.3mmol/L6
HDL-C10 – 110mg/dL0.26 – 2.84mmol/L6
IP0.5 – 15.0mg/dL0.16 – 4.84mmol/L5
Mg0.2 – 7.0mg/dL0.08 – 2.88mmol/L4.5
NH310 – 500μg /dL7 – 357μmol/L2
TBIL0.2 – 30.0mg/dL3 – 513μmol/L6
TCHO50 – 450mg/dL1.29 – 11.64mmol/L6
TCO25 – 40mmol/L5 – 40mmol/L5
TG10 – 500mg/dL0.11 – 5.65mmol/L4
TP2.0 – 11.0g/dL20 – 110g/L6
UA0.5 – 18.0mg/dL30 – 1071μmol/L4
ElectrolytesNa75 – 250mEq/L75 – 250mmol/L1
K1.0 – 14.0mEq/L1.0 – 14.0mmol/L
Cl50 – 175mEq/L50 – 175mmol/L
Immunological testCRP0.3 – 7.0mg/dL3 – 70mg/L5
Comprehensive S-panelTP, ALB, ALP, GLU, TBIL, IP, TCHO, GGT, GPT/ALT, Ca, CRE, BUN
Small Screening S-panelALP, TBIL, GGT, GPT/ALT, CRE, GOT/AST
  • * There are parameters which may not be available in your area. For details please contact your local distributor.
  • *1 Unit (A) and (B) is available

Calculations

Calculated Parameter

Indication

Unit

Equation

LDL Cholesterol

LDL

mg/dL

LDL-C = TCHO value - (HDL-C vlaue + TG value/5)

mmol/L

LDL-C = TCHO value - (HDL-C value + TG value/2.2)

non-HDL Cholesterol

non-HDL

mg/dL or mmol/L

non-HDL = TCHO value - HDL-C value

Globulin

GLOB

g/dL or g/L

GLOB = TP value - ALB value

Albumin/Globulin ratio

ALB/GLOB

-

ALB/GLOB = ALB value / (TP value - ALB value)

BUN/Creatinine ratio

BUN/CRE

-

BUN/CRE = BUN value / CRE value

Anion Gap

Anion Gap

mEq/L or mmol/L

Anion Gap = Na value - (Cl value + TCO2 value)

 

Main specifications

Measurement test

Colorimetry 28 tests
Electrolytes 3 tests

Throughput

Colorimetry 120 test/hour
Combined 128 test/hour

Number of sample rack

1

Number of incubator cell

Colorimetry 12, Electrolytes 1

Measurement time

Colorimetry 2 to 6 minutes/test,
Electrolytes 1 minute/3 tests (Na-K-Cl)

Sample type

Plasma, Serum, Whole blood*2

Sample volume

Colorimetry 10μL/test,
Electrolytes 50μL/3 tests (Na-K-Cl), CRP 5μL/test

Data transmission to PC

USB 2.0 or RS-232C Serial
D-Sub 9 pin -9 pin cross cable

Data print

Thermal Printer

Electrical requirements

AC 100-240V, 50/60Hz, 2.5-1.1A

Dimensions

470 (W) 5 360 (D) 5 420 (H) mm

Weight

NX500 Approx. 25kg, NX500i Approx. 24kg

Operating temperature

15 to 32˚C (59 to 89F)

Operating humidity

30 to 80%RH

  • *2 NH3-W: Whole blood only    NH3-P: Plasma only
  •  Na-K-Cl: Plasma, Serum, Whole blood
  •  Other test items: Plasma, Serum

DRI-CHEM  NX500 Series

 NX500NX500i
Electrolyte tests
Plasma Filter Function-
Automatic dilution
  • *DRI-CHEM NX500(Product:FUJI DRI-CHEM NX500/FUJI DRI-CHEM NX500i)
  • Please contact your local distributor for availability.

Option Items

Barcode reader

For reading sample ID on sample tube and operator ID

Operator ID Editor Kit

Operator ID Editor tool software (for PC) and a USB stick (for data transfer between PC and NX500) for registration and editing operator ID