Select Your Location

This is Fujifilm Portugal website.
To browse products and services available in your area,
please choose another country or region.

See all countries and regions

You are accessing from the United States. To browse Fujifilm USA website, please click the following link.

Fujifilm USA Website

DRI-CHEM IMMUNO AU10V - Specifications

A groundbreaking point-of-care immunological analyzer tuned for veterinary use. 

Overview

The content on this page is intended to healthcare professionals and equivalents.

Analyzer Specifications

Dimensions (W x D x H)

238 x 326 x 310 [mm]

Weight

Approximately 7 kg

Operating temperature

15 °C - 30 °C( 59 °F - 86 °F)

Setup time

3 min

Data memory

100 tests

Required sample volume

basic test : 100 μL

dilution test (v-COR): 20 μL

dilution test (vf-SAA): 10 μL

Cartridge information

Information about test parameters is embedded in the QR code attached to the cartridge.

Reagent Cartridges Specifications

 vf-SAAv-T4vc-TSHv-CORv-BAv-PRG
Target speciesFelineCanine / FelineCanineCanineCanine / FelineCanine
Type of sampleP / SP / SP / SP / SP / SP / S
Measurement Mode
(Basic)
✔ P / S✔ P / S✔ S only✔ P / S✔ P / S
Measurement Mode
(Dilution*1)
✔ P / S✔ P / S
DynamicRange3.75 - 225.0 μg/mL0.50 - 8.00 μg/dL
(6.4 - 103.0 nmol/L)
0.25 - 5.00 ng/mL1.0 - 30.0 μg/dL
(27.6 - 828.0 nmol/L)
1.0 - 50.0 μg/dL
(27.6 - 1380.0 nmol/L)
(with dilution)
2.0 - 150.0 μmol/L0.20 - 40.00 ng/mL
(0.64 - 127.20 nmol/L)
Measurement timeApproximately 10 min
Package Unit10 cartridges/box
Reagent Strorage2 °C – 8 °C (35.6 °F – 46.4 °F)

 

  • P : Plasma S: Serum
  • *1 Dilution with pharmaceutical grade saline solution

Dimensions

System Configuration Diagram

Options

Sample barcode reader