The content on this page is intended to healthcare professionals and equivalents.
Product Features
- Uses Silver Amplification technology for high detection sensitivity
- Time to results: Approximately 15 minutes
- Compatible with both visual and reader detection
- Nasopharyngeal swab
Silver Amplification technology
The immunochromatographic assay was designed based on silver halide photography to amplify the size of gold colloidal particles approximately a 100-fold, leading to the improvement of detection sensitivity. This offers a greater ability to detect even in small amounts of antigens.
How to use
Performance results
Clinical performance was established using retrospectively collected 225 samples (nasopharyngeal swabs).
Sensitivity: 90.9% (110/121, 95% CI: 84.3-95.4%)
Specificity: 100.0% (104/104, 95% CI: 96.5-100.0%)
RT-PCR | ||||
---|---|---|---|---|
Positive | Negative | Total | ||
FUJIFILM COVID-19 Ag Test | Positive | 110 | 0 | 110 |
Negative | 11 | 104 | 115 | |
Total | 121 | 104 | 225 |
Sensitivity according to the RT-PCR Ct value
RT-PCR Ct value | |||||
---|---|---|---|---|---|
< 26 | < 28 | < 30 | < 35 | ||
FUJIFILM COVID-19 Ag Test |
Positive | 100 | 107 | 110 | 110 |
Negative | 0 | 3 | 6 | 11 | |
Total | 100 | 110 | 116 | 121 | |
Sensitivity | 100.0% | 97.3% | 94.8% | 90.9% | |
95% CI | 96.4-100.0% | 92.2-99.4% | 89.1-98.1% | 84.3-95.4% |
Kit contents
FUJIFILM COVID-19 Ag Test (10 tests)
- Test Cartridge × 10 tests
- Extraction Reagent Solution R1 × 10 × 0.4mL
- Sterile Swab × 10 pieces
- Tube Rack (for Extraction Reagent Solution R1) × 1 piece
- Filter (for Extraction Reagent Solution R1) × 10 pieces
- Package Insert × 1 piece
Additional instrument (Optional, not included)
- FUJIFILM SILVAMP Reader