DRI-CHEM IMMUNO AU10V - Specifications

A groundbreaking point-of-care immunological analyzer tuned for veterinary use. 

Overview

The content on this page is intended to healthcare professionals and equivalents.

Analyzer Specifications

Dimensions (W x D x H)

238 x 326 x 310 [mm]

Weight

Approximately 7 kg

Operating temperature

15 °C - 30 °C( 59 °F - 86 °F)

Setup time

3 min

Data memory

100 tests

Required sample volume

basic test : 100 μL

dilution test (v-COR): 20 μL

dilution test (vf-SAA): 10 μL

Cartridge information

Information about test parameters is embedded in the QR code attached to the cartridge.

Reagent Cartridges Specifications

 vf-SAAv-T4vc-TSHv-CORv-BAv-PRG
Target speciesFelineCanine / FelineCanineCanineCanine / FelineCanine
Type of sampleP / SP / SP / SP / SP / SP / S
Measurement Mode
(Basic)
✔ P / S✔ P / S✔ S only✔ P / S✔ P / S
Measurement Mode
(Dilution*1)
✔ P / S✔ P / S
DynamicRange3.75 - 225.0 μg/mL0.50 - 8.00 μg/dL
(6.4 - 103.0 nmol/L)
0.25 - 5.00 ng/mL1.0 - 30.0 μg/dL
(27.6 - 828.0 nmol/L)
1.0 - 50.0 μg/dL
(27.6 - 1380.0 nmol/L)
(with dilution)
2.0 - 150.0 μmol/L0.20 - 40.00 ng/mL
(0.64 - 127.20 nmol/L)
Measurement timeApproximately 10 min
Package Unit10 cartridges/box
Reagent Strorage2 °C – 8 °C (35.6 °F – 46.4 °F)

 

  • P : Plasma S: Serum
  • *1 Dilution with pharmaceutical grade saline solution

Dimensions

System Configuration Diagram

Options

Sample barcode reader