Clinical biochemistry, enzyme, immunological and electrolyte analytes
Fujifilm Dri-Chem has evolved its dry biochemistry technology over the past four decades, optimizing systems for simplified, intuitive operation with high accuracy.
- Automated System
- Maximum of five samples in one operation
- Integrated touchscreen and intuitive interface
- No calibration required*1 Simple batch compensation via QC card
- *1 except for CRP
Benchtop equipment, no structural modifications required.
- Automatic dual voltage: 100 to 240v
- Does not use water in operation
- Does not require air supply or exhaust
Product dimensions
| Width 500 mm | Weight of main unit 33 kg |
| Depth 380 mm | |
| Height 410 mm |
The sliding lid avoids unnecessary space consumption and facilitates handling. The ease of use of the new design allows for smoother operation. The exterior was specifically designed to improve adaptability to any location.
Simple and easy-to-use operating screen. Explanatory diagrams facilitate use, troubleshooting, and maintenance procedures.
![[photo] FUJI DRI-CHEM NX700 with lid open, exposing internal working parts; ticket of results printing out of top of machine](https://asset.fujifilm.com/www/br/files/2020-05/893a157190217c97e7fe3d2349d9b14b/pic_dri-chem-nx700_03.jpg)
Setup screen
Display of results
Help screen
Set the slide, the specimen and press the START key. All the processes hereafter are fully automated.
A maximum of 5 specimens can be set at the same time. No manual operation after pressing START key. The automation shortens operation time and improves workflow efficiency.
Batch variation corrections are made simply by reading the QC card included in the blade package.
Each test requires only 10 μL of sample. (The CRP requires 5 μL/test, the ISE requires 50 μL/3 tests). Manual pipetting can also be performed when the sample is smaller. Less invasive for newborns and patients requiring high-complexity care.
- * 50 μL are used for the simultaneous measurement of 3 electrolyte items.
After measurement, the pipette tips and slide reagents are automatically discarded into the disposal box, minimizing the risk of contamination.
In addition to Fuji tubes (0.5/1.5 mL), commercially available blood collection tubes can be used as specimen tubes.
- * Please select suitable sample racks upon your work flow. (refer last page:Option items/Sample Rack)
Set CRP slides, diluent and dedicated calibrator into the specimen disk. Easy calibration starts by pressing the “Calibration” key.
Labor intensive operations like dispensing, mixing etc. are automated. The only operation is to input the dilution ratio.
Electrolytes (Na-K-Cl) can also be measured. With ISE module technology, the gold standard for electrolytic testing.
Press the “STAT” key when there is ongoing measurement. Set the emergency specimen and just press START to perform emergency test.
The NX700 can communicate using standard HL7 protocol.
Identification of operators can be set.
Quality control in the NX700 is made easier with the advance QC functions. Results of quality control can be viewed on the NX700 screen.
![[image] Turquoise, black, and grey setup screen on incubator's display with Operator ID](https://asset.fujifilm.com/www/br/files/2020-05/85f505683e027f7ad3481e7c233d7ad1/pic_dri-chem-nx700_14.jpg)
![[image] Control fluid setting screen/setup for quality control function](https://asset.fujifilm.com/www/br/files/2020-05/5f5439e35320952fa59a8b41fcf4caed/pic_dri-chem-nx700_15.jpg)
![[image] Dark control chart screen with list of dates and results plotted on chart](https://asset.fujifilm.com/www/br/files/2020-05/1a6c93a0e3e6bc60b6c8d4cddb53f1cb/pic_dri-chem-nx700_16.jpg)
FUJI DRI-CHEM slide reagent boasts high reliability and stability, backed by dry chemical technology cultivated throughout FUJIFILM's long history in the photographic film manufacturing industry. It offers less variation in results between operators, high reproducibility and daily accuracy, and excellent correlation.
![[image] GOT/AST chart of FUJI DRI-CHEM Slide reagent and IFCC standardization method](https://asset.fujifilm.com/www/br/files/2020-05/ee5fa767bd2144f6c0d9703a11384635/fig_dri-chem-nx700_01.jpg)
![[image] GLU chart of FUJI DRI-CHEM Slide reagent and Hexokinase method](https://asset.fujifilm.com/www/br/files/2020-05/c6fbe1307dfb3dd9cf67ebd2ec90e7e6/fig_dri-chem-nx700_02.jpg)
![[image] K chart of FUJI DRI-CHEM Slide reagent and Ion-selective electrode method/dilution method](https://asset.fujifilm.com/www/br/files/2020-05/2403b92116cb019db037604b41b36f0d/fig_dri-chem-nx700_03.jpg)
![[image] CRP chart of FUJI DRI-CHEM Slide reagent and Latex immunoturbidimetry](https://asset.fujifilm.com/www/br/files/2020-05/225494a6c0ffea88c3ba3456b240126f/fig_dri-chem-nx700_04.jpg)
(Enzymes, General chemistry, and Immunology)
This multilayered slide is composed of dry chemical ingredients needed for the reaction and other functional materials. It quantifies enzymes and chemicals using colorimetric method.
Composition of multilayered analytical film
(Electrolytes)
The electrolytes slide contains ion selective film electrodes for Na, K, and Cl. These electrolytes are measured by potentiometric method.
Composition of multilayered film electrode
| Series name | FUJI DRI-CHEM NX700 | FUJI DRI-CHEM NX500 |
| Width × depth × height | 500(W) × 380(D) × 410(H) mm | 470(W) × 360(D) × 420(H) mm |
| Weight | 33 kg | 25 kg |
| Number of measurement cells | 13 colorimetric + 1 electrolyte (independent) | 12 colorimetric + 1 electrolyte (independent) |
| Throughput (1) (only colorimetry) | 180 tests/hour | 120 tests/hour |
| Throughput (2) (colorimetry + electrolytes) | 190 tests/hour | 128 tests/hour |
| Throughput (3) (15-item measurements) | approx. 9 minutes | approx. 9 minutes |
| Number of specimen that can be set | 5 specimens at the same time | 1 specimen |
| Maximum memory of QC information/ parameter | 5 lots | 2 lots |